built-in inside the bottom of Cellbag), the concentration of O2 and
CO2 in the inlet gas mixture, as well as the total gas flow rate. The
range of operating parameters that can be applied in system of
WAVE 25 has been presented in Table 1.
Fig. 2 Schematic diagram of Cellbag container with arrangement of inlet and
outlet ports: (1) Cellbag rods, (2, 6, 7) various pipe-ports, (3) outlet vent filter with
pressure control valve, (4) inlet vent filter, (5) Screwcap port, (8) CLAVE™
sampling port
Table 1
The range of operating parameters that can be applied in WAVE 25 system
Operating parameters
Range of parameter
Unit
Angle of oscillations
2–12
Frequency of oscillations
2–40
min1
Gas flow rate
0.1–1.0
L min1
Volume of liquid phase
0.1–1.0
L
Temperature
20–40
C
O2 concentration in gas phase
0–50
%
CO2 concentration in gas phase
0–15
%
Rocking motion acceleration
15–100
%
150